• Home
  • Products
  • Order
  • Support
  • Contact

Troubleshooting - Suspended Solids Monitor Model 950


Download print version of troubleshooting information

Problem Possible Cause Check/Remedy
Meter indication is blank • No power
• Blown fuse
• No power to instrument
• Faulty transmitter/indicating meter
• Check power light.
• Check fuse.
• Check for proper voltage on terminals 8 & 9.
• Refer to supplier.
Meter will not zero (high reading) • Too much absorbed light


• Poor sensor connections or incorrect wiring
• Sensor leads not insulated from body of sensor or conduit
• Wet connection in converter or sensor housings
• No power to sensor



• Faulty sensor bulb or defective sensor/transmitter
• Remove sensor and place in clean water. If the meter does not zero then try adjusting CS (course zero) on the circuit board.
• Check connections and wiring.

• Check sensor wires.


• Look for water or condensate on connections.

• Voltage should be 7-10 VDC between terminals 5 and 6 on the main circuit board. Also, there should be 6-9 VDC on terminals 5 and 6 of the circuit board in the sensor.
• If the proper voltage (8-11 VDC) is on terminals 5 and 6 and the bulb in the sensor is not on, then the bulb has failed.
No response (indication does not change and indication is high) • Faulty sensor bulb • Check to see if sensor lamp is on. If lamp is off, check voltage across terminals 5 & 6 in sensor. If the proper voltage (8-11 VDC) is on terminals 5 and 6 and the bulb in the sensor is not on, then the bulb has failed.
Unstable reading • Insufficient dampening
• Air bubbles or very large particles in process line
• RFI pickup
• Heavy inductive loads
• Increase dampening.
• Take sample and check for bubbles or large particles. If present, relocate sensor.
• Earth ground instrument properly.
• Mount instrument away from power cables. Put cables in conduit.
Reading does not agree with lab results • Improper calibration
• Lab procedure error
• Lab instrument error
• Defective sensor
• Recalibrate instrument.
• Check procedure.
• Check instrument.
• Refer to supplier.
Readings drift with time • Converter not warmed up
• Sensor/converter connections wet

• Deposit buildup on sensor
• Warm up converter for 15 minutes.

• Look for water or condensate on connections. Dry connections with hair dryer.

• Remove sensor and clean.
Negative readings or negative output • Improper calibration • Recalibrate instrument.

Turbidimeter

Turbidimeter Model 850
  • Description
  • Specifications
  • Sensor Mounting
  • Pricing

Suspended Solids Monitor

Suspended Solids Monitor Model 950
  • Description
  • Specifications
  • Sensor Mounting
  • Pricing

Copyright © 2009 | Contact