Confab logo

Turbidimeter •  Product Description
•  Specifications
•  Sensor Mounting
•  Pricing


Suspended Solids Monitor •  Product Description
•  Specifications
•  Sensor Mounting
•  Pricing


More Info •  Information Request
•  Email Webmaster
•  Tech Support
 

If you have technical questions on our products, please refer to the appropriate troubleshooting guide below (Model 850 or Model 950). If you do not find the answer to your question here, please refer to our detailed installation and instruction manuals:

  • Download Model 850 Installation and Instruction Manual (PDF, 481K)
  • Download Model 950 Installation and Instruction Manual (PDF, 565K)

    Troubleshooting for Model 850 Turbidimeter

    Problem

    Possible Causes

    Check/Remedy

    No response/meter reading blank

    No power

    Blown fuse

    Faulty sensor bulb

    Check power light.

    Check fuse.

    Check to see if sensor lamp is on. If lamp is off, check voltage across terminals 5 & 6 in sensor. If voltage is greater than 7 VDC, then sensor has failed.

    Meter will not zero (high reading)

    Too much scattered light

     

    Too much ambient light or too much reflected light

     



    Poor sensor connections or incorrect wiring

    Sensor leads not insulated from body of sensor or conduit

    Wet connection in converter or sensor housings

    Defective sensor/transmitter

    Remove sensor and place in clean water with a cover to shield ambient light.

    Disconnect wires on terminals 5 & 6. If meter still will not zero then ambient light is entering pipe. If meter will zero, then too much reflected light. Try repositioning the sensor to reduce reflection.

    Check connections and wiring.

    Check sensor wires.


    Look for water or condensate on connections.

    Remove all sensor wires at transmitter. If meter will not zero turn span full counter clockwise. If meter still will not zero, transmitter is bad. If meter will zero then sensor possibly has failed.

    Unstable reading

    Air bubbles or very large particles in process line


    RFI pickup


    Heavy Inductive loads

    Take sample and check for bubbles or large particles. If present relocate sensor.

    Earth ground instrument properly.

    Mount instrument away from power cables. Put cables in conduit.

    Reading does not agree with lab results

    Improper calibration

    Lab procedure error

    Lab instrument error

    Defective sensor

    Recalibrate instrument.

    Check procedure.

    Check instrument.

    Refer to supplier.

    Readings drift with time

    Converter not warmed up


    Sensor/converter connections wet




    Deposit buildup on sensor

    Warm up converter for 15 minutes.

    Look for water or condensate on connections. Dry connections with hair dryer. Then use NEMA 4X enclosure.

    Remove sensor and clean.

    Negative readings or negative output

    Wet or damp connections



    Faulty sensor bulb

    Dry connections with hair dryer. Then use NEMA 4X enclosure.

    Check to see if sensor lamp is on. If lamp is off, check voltage across terminals 5 & 6 in sensor. If voltage is greater than 7 VDC, then sensor has failed.

    Troubleshooting for Model 950 Suspended Solids Monitor

    Problem

    Possible Causes

    Check/Remedy

    Meter indication is blank

    No power

    Blown fuse

    No power to instrument

    Faulty transmitter/indicating meter

    Check power light.

    Check fuse.

    Check for proper voltage on terminals 8 & 9.

    Refer to supplier.

    Meter will not zero

     

     

     

     

    Too much absorbed light





    Poor sensor connections or incorrect wiring

    Sensor leads not insulated from body of sensor or conduit
    Wet connection in converter or sensor housings

    No power to sensor






    Faulty sensor bulb or defective sensor/transmitter

    Remove sensor and place in clean water. If the meter does not zero then try adjusting CS(course zero) on the circuit board.

    Check connections and wiring.

    Check sensor wires.


    Look for water or condensate on connections.

    Voltage should be 7-10 VDC between terminals 5 and 6 on the main circuit board. Also, there should be 6-9 VDC on terminals 5 and 6 of the circuit board in the sensor.

    If the proper voltage (8 to 11VDC) is on terminals 5 and 6 and the bulb in the sensor is not on, then the bulb has failed.

    No response (indication does not change and indication is high)

    Faulty sensor bulb

    Check to see if sensor lamp is on. If lamp is off, check voltage across terminals 5 & 6 in sensor. If the proper voltage (8 to 11VDC) is on terminals 5 and 6 and the bulb in the sensor is not on, then the bulb has failed.

    Unstable reading

    Insufficient dampening

    Air bubbles or very large particles in process line


    RFI pickup


    Heavy inductive loads

    Increase dampening.

    Take sample and check for bubbles or large particles. If present, relocate sensor.

    Earth ground instrument properly.

    Mount instrument away from power cables. Put cables in conduit.

    Reading does not agree with lab results

     

    Improper calibration

    Lab procedure error

    Lab instrument error

    Defective sensor/transmitter

    Recalibrate instrument.

    Check procedure.

    Check instrument.

    Refer to supplier.

    Readings drift with time

     

    Converter not warmed up


    Sensor/converter connections wet


    Deposit build-up on sensor

    Warm up converter for 15 minutes.

    Look for water or condensate on connections. Dry connections with hair dryer.

    Remove sensor and clean.

    Negative readings or negative output

    Improper calibration

    Recalibrate instrument.


    © 2006   Confab Instrumentation • 11180 Ranchette Drive • Jackson CA 95642 • 209 223 1090